• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RETROGRADE DRL 9MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RETROGRADE DRL 9MM; BIT, SURGICAL Back to Search Results
Model Number 72204044
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Information (3190)
Date of Event 07/16/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an acl reconstruction, when using the trunav retrograde drill 9mm in the femoral cortex, it was found that the tip was detached.The tip of the device was removed from the patient.The procedure was completed without delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
One 72204044 retrograde drill used in treatment, has not retuned for evaluation.Without the reported product, a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.However the customer has provided an image.The information provided states: ¿during acl reconstruction, when using the trunav retrograde drill 9mm in the femoral cortex, it was found that the tip was detached¿.Visual assessment of image confirms breakage.An exact root cause cannot be determined with confidence.The instruction for use was reviewed and was found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.Complaint history review indicated a similar allegation for the lot number reported.Batch review indicated no condition, product or procedure failure that supported the allegation.Product met specifications upon release to distribution.Occurrence rate of allegations are monitored via surveillance.No further investigation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 9MM
Common Device Name
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10383249
Report Number1219602-2020-01212
Device Sequence Number11498560
Product Code GFG
UDI-Device Identifier00885554037739
UDI-Public00885554037739
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2024
Device Model Number72204044
Device Catalogue Number72204044
Device Lot Number4899147
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer10/14/2020
Initial Report FDA Received Date08/09/2020
Supplement Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-