| Model Number |
MMT-1715KL |
| Medical Device Problem Codes |
Moisture or Humidity Problem (2986); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had critical pump error.Customer reported they received the open book image on the insulin pump screen and exposed to moisture.Customer stated no insulin pump error displayed on the screen before the open book image.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with critical pump error due to moisture damage to force sensor.Device received with corroded electronic assemblies and corroded motor home switch.Unable to perform displacement test, rewind test, prime or seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self test due to critical pump error alarm noted.
|
| |
|