| Model Number |
MMT-7020A |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 91mg/dl and the sensor glucose was 40 mg/dl.Insulin delivery was suspend due to sensor values.Customer stated insulin pump received calibration not accepted and change senor alert.The sensor will not be returned for analysis.
|
| |
|