| Model Number |
MMT-1715KL |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received black display after replacing battery.Customer removed battery and received insert battery alarm.Customer checked if battery cap contacts were damaged or missing.Customer checked if battery compartment and spring were damaged or corroded.Customer insert battery but display did not return.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement, sleep current measurement, active current measurement, and self tests.No unexpected blank displays or unexpected ¿low battery¿, ¿replace battery now¿, or "power loss" errors were noted during testing.(b)(4).
|
| |
|