| Model Number |
SCM-175 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
07/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "infection¿ is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: infection (late onset).
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side infection.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.' further investigation results: with the information collected during the investigation, there is enough evidence to support that device serial number (b)(6) was manufactured under controlled conditions in accordance with allergan procedures.Seal strength results were acceptable.Environmental monitoring results were found to be acceptable and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run (b)(4) is not related to any additional complaint infection record reported as of today.During the trend review of all infection complaints for gel breast implants for the period of (b)(6) 2018 through (b)(6) 2020, were noted 2 outliers ((b)(6) 2018 and (b)(6) 2019).An additional analysis was performed to determine any pattern or any similarities relation between prs involved.It was found a primary packaging operator involved in more than one occasion in this process, however the person was trained in the corresponding procedure.Also, it was found a sealer and a sterilizer involved in more than one occasion on those processes, however, calibrations, maintenance and the validation processes of these equipment involved, were reviewed and it was determined that they were performed on time and met specifications.In addition, all the records involved in this month were reviewed and no issues were found associated to either event.Given the fact that the cause of the infection cannot be specifically associated to the manufacturing process, and there is not an adverse trend for this type of event no corrective action is deemed necessary at this time.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side infection.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|