Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Correction: section c: the event occurred in the (b)(6).Summary of event: as reported, during an angioplasty an advance 35 lp low profile balloon catheter was in use when it burst after it had been inflated to 8 atm.It is unknown how the procedure was completed.No portion of the device remained inside the patient.This did not result in any additional procedures or prolonged hospitalization.No adverse effects to the patient occurred.Investigation evaluation: reviews of the complaint history, device history record, drawing, documentation, instructions for use (ifu), manufacturing instructions, quality control, and specifications were conducted during the investigation.A visual inspection of the complaint device was also conducted.The complaint device was returned to cook for investigation with biomatter present.Physical examination of the returned device showed that there was no visible damage to the catheter shaft.The balloon was inflated using water and a pinhole leak was detected at 1cm from the proximal bond, confirming the reported failure.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The information provided upon review of complaint file, device history record, complaint history, device master record, and design verification testing provide objective evidence to support that the device was manufactured to specification.As there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other lot related complaints have been received from the field, it was concluded that there is no evidence that nonconforming product exists in-house or in the field.The label on the product package and the compliance card insert indicate that for the (b)(6), the nominal inflation pressure is 8atm and the rated burst pressure is 11atm.The ifu also instructs the user to adhere to the balloon inflation pressure parameters indicated on the labeling.The ifu warns not to exceed the rated burst pressure, as rupture may occur.A device master record review was performed, including device specifications, drawings, manufacturing instructions, and quality control procedures.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that a definitive cause for the event cannot be determined.The risk analysis for this failure mode was reviewed and no action was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|