• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS Back to Search Results
Model Number 315-0C-717
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/14/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as infection.The previous surgery and the surgery detailed in this event occurred 28 days apart.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was kept at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material report (ncmr) associated with the products that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was reported as an infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the possible infection or inhibited the patient's immune system.Due to the short time between the original surgery and the revision, it is possible that the infection was acquired in the hospital (nosocomial).It is also possible that the patient was not compliant with post surgical instructions.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Event or Problem Description
Revision surgery - due to infection and the poly was swapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT REVISION KNEE
Common Device Name
INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key10384676
Report Number1644408-2020-00661
Device Sequence Number2898509
Product Code JWH
UDI-Device Identifier00888912114103
UDI-Public(01)00888912114103
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/04/2024
Device Model Number315-0C-717
Device Catalogue Number315-0C-717
Device Lot Number013R1015
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2020
Initial Report FDA Received Date08/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-