MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Poor Quality Image (1408)
Patient Problem No Patient Involvement (2645)
Event Date 03/09/2020
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation as it functions as intended.In addition, tac suggested that the customer replace the video cable as the phenomenon can reoccur.
 
Event Description
The service center was informed that a green tint was noted on the device image.The technical assistance center (tac) provided troubleshooting assistance and instructed the customer to reseat the video cable.The cable was reseated and the image issue was resolved.There was no patient involvement reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10384813
MDR Text Key202207938
Report Number8010047-2020-05406
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/13/2020
Initial Date FDA Received08/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N