| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Codes |
Head Injury (1879); Failure of Implant (1924)
|
| Date of Event |
07/10/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user's parents reported an obvious decline in hearing performance with the device after a head trauma.After three weeks of observation, his hearing didn't improve.The user has been re-implanted on (b)(6) 2020.
|
| |
|
Event or Problem Description
|
|
The user_s parents reported an obvious decline in hearing performance with the device after a head trauma.After three weeks of observation, his hearing didn't improve.The user has been re-implanted on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: the investigation results confirmed that the device has failed due to an external impact to the active electrode.All the problems given in the recipient report appear to match well with this finding.This is a final report.
|
| |
|