| Model Number |
CD3265-40 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented remotely via merlin.Net.Review of the transmission revealed non-sustained right ventricular oversensing due to post paced t wave oversensing.The oversensing was noted on a stored egm (electrograms).The patient did not receive therapy during the oversensing.Programming changes were made.There were no patient consequences.
|
| |
|