| Model Number |
7205396 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is under investigation by the manufacturing site.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure the footswitch was broken.No delay and a back up was available to complete the procedure.No patient injury was reported.Preliminary results of investigation have concluded that both pedals of the footswitch where stuck in the depressed position which makes it a reportable event.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection of the returned device noted that the both pedals are stuck in the depressed position.Functional evaluation revealed a footswitch error message due to the stuck pedals.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.The complaint was verified.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.
|
| |
|