Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.However, device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, self test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test or verify no excessive no delivery or occlusion alarm due to motor error alarms.Device received with cracked case at the reservoir tube window corners, cracked reservoir tube window, broken battery tube threads and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|