MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAP; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAP
Medical Device Problem Codes Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump bolus button harder to press.Customer reported that the insulin pump rejected new batteries and unexplained no insulin delivery alert.Customer reported that the insulin pump had squirts on prime, crack on reservoir.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device received with intermittent button response due to corroded keypad traces.No button error alarm during testing.However, device alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, self test, rewind, basic occlusion test, occlusion test, prime/a33 test and excessive no delivery test or verify no excessive no delivery or occlusion alarm due to motor error alarms.Device received with cracked case at the reservoir tube window corners, cracked reservoir tube window, broken battery tube threads and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAP
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10389318
Report Number2032227-2020-160650
Device Sequence Number16613100
Product Code OYC
UDI-Device Identifier00613994402103
UDI-Public(01)00613994402103
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAP
Device Catalogue NumberMMT-522NAP
Device Lot NumberA5522NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2020
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer10/09/2020
Initial Report FDA Received Date08/11/2020
Supplement Report FDA Received Date10/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight133
-
-