MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 670G INSULIN PUMP MMT-1780KL; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-1780KL
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump would not turn on.No harm requiring medical intervention was reported.The device will not be returned for analysis.
 
Additional Manufacturer Narrative
The device p-cap / test reservoir locks in place properly.The insulin pump powered up properly after battery installation.No blank display noted.The insulin pump passed the displacement test, sleep current measurement, active current measurement and self test.The insulin pump was monitored and no unexpected powering off or blank display noted.The insulin pump was cut open and verified the battery tube power connector is properly connected to electrical board 1 during visual inspection.No moisture on electronic assembly noted.The insulin pump was received with a cracked keypad overlay, cracked case (battery tube) and cracked battery tube threads.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
670G INSULIN PUMP MMT-1780KL
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10389425
Report Number2032227-2020-160721
Device Sequence Number2259542
Product Code OZP
UDI-Device Identifier00763000365882
UDI-Public(01)00763000365882(17)230709
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/09/2023
Device Model NumberMMT-1780KL
Device Catalogue NumberMMT-1780KL
Device Lot NumberHG4G794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer09/02/2021
Initial Report FDA Received Date08/11/2020
Supplement Report FDA Received Date09/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age31 YR