The device p-cap / test reservoir locks in place properly.The insulin pump powered up properly after battery installation.No blank display noted.The insulin pump passed the displacement test, sleep current measurement, active current measurement and self test.The insulin pump was monitored and no unexpected powering off or blank display noted.The insulin pump was cut open and verified the battery tube power connector is properly connected to electrical board 1 during visual inspection.No moisture on electronic assembly noted.The insulin pump was received with a cracked keypad overlay, cracked case (battery tube) and cracked battery tube threads.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|