• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA PEDICLE SCREW 8X40MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA PEDICLE SCREW 8X40MM Back to Search Results
Model Number 03.50.054
Health Effect - Clinical Code Joint Disorder (2373)
Date of Event 07/20/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 30 july 2020: lot 1924057: (b)(4) items manufactured and released on 03-mar-2020.Expiration date: 2025-02-08.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported event.Other devices involved in the event pedicle screw 03.56.404 must straight cross connector 40mm lot.1720364 (k171170).Batch review performed on 30 july 2020 lot 1720364: (b)(4) items manufactured and released on 19-jul-2017.Expiration date: 2022-07-03.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold with two other similar reported events.
 
Event or Problem Description
Spinal fixtaion was performed at l2-l5 firstly.The surgeon engaged the pedicle screw with the screw driver, but he felt the engage strength of the pedicle screw and the screwdriver might be weak.After the insertion of the pedicle screw and the rod, the surgeon felt the loosening of the pedicle screw at the left side of the l5, so the surgeon replaced the pedicle screw and the rod and performed the additional spinal fixtaion.Moreover when the surgeon engaged the cross connector, the arm of the cross connector contacted to the lamina.In order to use it the surgeon performed additional lamina cut.The surgeon did not use the cross connector because the surgeon would like to avoid additional cut.Finally, l2-s1 was fixed.Due to this event the surgery was prolonged about 25 minutes.Total surgery time was about 150mins.
 
Additional Manufacturer Narrative
Visual inspection.During the visual inspection, it has been analized: the pedicle screw, ref.03.50.054, lot.1924057; the bent rod, ref.03.52.442, lot.1620911; the must straight cross connector, ref.03.56.404, lot.1720364; the polyaxial screwdriver reduction, re.03.51.10.0222, lot 1850161.All the implants analized didn't present any damaging or unfunctionality, except the must straight cross connector for which is missed a screw.The instrument analized didn't present any damaging or unfunctionality.The functionality between the pedicle screw and the polyaxial screwdriver reduction has been test and the fixation is stable.The must straight cross connector has small encumbrance but it shoulf be available as alternative the must link cross connector that can give to the surgeon addiotnal option to avoid lamina cut.Visual inspection performed on august the 31st, 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDICLE SCREW 8X40MM
Common Device Name
PEDICLE SCREW
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874, CHE
SZ  6874, CHE
MDR Report Key10389668
Report Number3005180920-2020-00506
Device Sequence Number2259352
Product Code NKB
UDI-Device Identifier07630030834417
UDI-Public07630030834417
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132878
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device No Information
Device Expiration Date02/08/2025
Device Model Number03.50.054
Device Catalogue Number03.50.054
Device Lot Number1924057
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2020
Supplement Date Received by Manufacturer07/20/2020
Initial Report FDA Received Date08/11/2020
Supplement Report FDA Received Date09/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-