|
|
| Model Number |
030418 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, pre operatively, on a laparoscopic left upper lobectomy, when the device was assembled, the stapler was unable to fire, the surgeon was able to press the green button and squeeze the handle.The instrument was replaced to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.Visual inspection noted that the returned device was pre-fired, engaged in interlock, and the jaws were clamped.It was reported that the device did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the shipping wedge was removed or no longer fully seated in the metal channel prior to loading, the shipping wedge was removed or no longer fully seated in the metal channel and the jaws were manually clamped prior to loading, or the shipping wedge was removed or no longer fully seated in the metal channel and an attempt was made to load a loading unit with the instrument firing rod extended.Please note that the yellow shipping wedge must be securely in place during instrument loading.The shipping wedge ensures that the single use loading unit (sulu) engages in the instrument to facilitate clamping and unclamping.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not attempt to remove the shipping wedge until the sulu is loaded into the instrument.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, pre operatively, on a laparoscopic left upper lobectomy, when the device was assembled, the stapler was unable to fire, the surgeon was able to press the green button and squeeze the handle.The staple was replaced to resolve the issue.There was no patient injury.
|
| |
|
Search Alerts/Recalls
|
|
|