MEDTRONIC IRELAND DEFLECT; CATHETER, PERCUTANEOUS Back to Search Results
Model Number C315HIS02
Device Problems Break (1069); Deformation Due to Compressive Stress (2889)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.Analysis indicated the mechanical operation of the catheter shaft was kinked/buckled.Blood was observed in the hemostasis valve of the delivery catheter.Blood was observed on the shaft of the delivery catheter.Visual analysis of the lead indicated damage during use.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, when the physician inserted the lead into the catheter and moved it to the target location, the lead seemed to get stuck.A kink in the catheter was found during fluoroscopy and the catheter was removed.Upon inspection, a deep kink was found and the catheter broke.A new catheter was used to implant the lead.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEFLECT
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
IE 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
IE  
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10390973
MDR Text Key202403399
Report Number9612164-2020-02951
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00613994749925
UDI-Public00613994749925
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K101885
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/05/2022
Device Model NumberC315HIS02
Device Catalogue NumberC315HIS02
Device Lot Number0010125835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/20/2020
Initial Date FDA Received08/11/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-