(b)(4).The device will not be returned for analysis as it was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00369, 0001032347-2020-00370, 0001032347-2020-00371.Concomitant medical products: sternalock 360 multi-implant system, part# 74-0004, lot# 065990.Sternalock blu system screw, cancellous cross-drive locking 2.4 x 14mm, part# 73-2414, lot# ni.Sternalock blu system screw, cancellous cross-drive locking 2.4 x 16mm, part# 73-2416, lot# ni.Sternalock blu system screw, cancellous cross-drive locking 2.4 x 18mm, part# 73-2418, lot# ni.
|
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record for part 74-0004 identified no deviations or anomalies during manufacturing related to the reported event.Review of the device history records identified no deviations or anomalies during manufacturing for the three plate subcomponents.The doctor reported the plate broke and screws ripped out due to the patient¿s cough; however, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|