| Model Number |
ZCU525 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.Date of event: unknown, not provided; best estimate is between (b)(6) 2020.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis toric ii intraocular lens (iol) was explanted from the patient's eye due to the need for new lens power for patient to reach the intended vision.The replacement lens is the same model, but higher diopter (22.5) iol.There were no sutures or incision enlargement required.Reportedly, patient is doing well post-operation.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: device available for evaluation: yes section d10: returned to manufacturer on: 8/21/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was returned in the replacement lens¿ lens case.The replacement lens¿ folding carton, patient stickers, and dfu were received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|