JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS TORIC II 1-PIECE ACRYLIC; TORIC IOLS Back to Search Results
Model Number ZCU525
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.Date of event: unknown, not provided; best estimate is between (b)(6) 2020.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis toric ii intraocular lens (iol) was explanted from the patient's eye due to the need for new lens power for patient to reach the intended vision.The replacement lens is the same model, but higher diopter (22.5) iol.There were no sutures or incision enlargement required.Reportedly, patient is doing well post-operation.No further information provided.
 
Additional Manufacturer Narrative
Section d10: device available for evaluation: yes section d10: returned to manufacturer on: 8/21/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was returned in the replacement lens¿ lens case.The replacement lens¿ folding carton, patient stickers, and dfu were received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS TORIC II 1-PIECE ACRYLIC
Common Device Name
TORIC IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10393806
Report Number9614546-2020-00301
Device Sequence Number16847486
Product Code HQL
UDI-Device Identifier05050474664319
UDI-Public(01)05050474664319(17)241021
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/21/2024
Device Model NumberZCU525
Device Catalogue NumberZCU525U215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/16/2020
Supplement Date Received by Manufacturer07/16/2020
10/25/2020
Initial Report FDA Received Date08/11/2020
Supplement Report FDA Received Date09/15/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-