EXACTECH, INC. LOGIC TIBIA IMPLANT PSC INSERT, SZ 2, 13MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
|
Back to Search Results |
|
| Model Number |
02-012-44-2013 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2016, a (b)(6) y/o, female patient with a bmi of 25.7, underwent implantation of a insert tib 2 13mm knee post stab cnstrn mod net cmpr mld.On (b)(6) 2018, approximately 23 months post implant, the patient underwent revision due to aseptic loosening.
|
| |
|
Additional Manufacturer Narrative
|
|
(h3) the evaluation noted that revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening.However, this cannot be confirmed as the devices were not returned for evaluation.
|
| |
|