LEONHARD LANG GMBH STABLE BASE; ECG ELECTRODE Back to Search Results
Model Number T-601
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Irritation (1941); Itching Sensation (1943); Rash (2033); Burning Sensation (2146)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Retained samples of the concerned lot/batch of model sbt601 have been inspected visually.Mechanical tests were performed on 3 retained samples.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect and details of the use was provided.We have requested further information.We will relay any further information in a follow up report.
 
Event or Problem Description
On july 28th, 2020, we have been informed about 9 incidents with ecg electrodes.Monitoring ecg electrodes (model sbw601 and sbt601) had been used with bodyguardian mini+ and bodyguardian one devices.The initial reporter has attached for each patient an incident summary.In total, there were 9 reports covering a period from (b)(6).No medical intervention was necessary for 8 incidents.For 1 incident a medical intervention was needed to treat the injury.The reportable event disclosed that the patient's skin was prepared using water and soap before applying the ecg electrodes.The patient experienced the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." it was also reported that the patient stopped the study after 5 days.It was stated that a medical intervention was needed treat the injury but no details were given.No further information was provided.
 
Additional Manufacturer Narrative
Retained samples of the concerned lot/batch of model sbt601 have been inspected visually.Mechanical tests were performed on 3 retained samples.All tested samples were found to perform within limits.No faults could be detected.No information on the patient, skin type, state of skin, whether any medication was being taken, which might have a skin weakening effect, duration of use and details of the use was provided.We have requested further information several times but none was made available.The initial reporter was specifying "we continue to encourage the customer to provide this information.However we can't force them to do so.Our customer is a very large healthcare provider in the united states and they have a general system for patient complaints.They will not develop a special tracking system and process for leonhard lang." after requesting for several information on the usage and wearing durations we have been informed that: "we have spent years trying to get this provider to add additional questions to their patient complain process ( forms ) with no luck.I'm not sure there is much more i can do at this time." we therefore consider the investigation closed.
 
Event or Problem Description
On (b)(6) 2020, we have been informed about 9 incidents with ecg electrodes.Monitoring ecg electrodes (model sbw601 and sbt601) had been used with bodyguardian mini+ and bodyguardian one devices.The initial reporter has attached for each patient an incident summary.In total, there were 9 reports covering a period from (b)(6).No medical intervention was necessary for 8 incidents.For 1 incident a medical intervention was needed to treat the injury.The reportable event disclosed that the patient's skin was prepered using water and soap before applying the ecg electrodes.The patient experienced the "redness, itching, rash, blisters, welt, burning sensation on skin, peeling skin, burnt skin, bleeding, open sores." it was also reported that the patient stopped the study after 5 days.It was stated that a medical intervention was needed treat the injury but no details were given.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STABLE BASE
Common Device Name
ECG ELECTRODE
Manufacturer (Section D)
LEONHARD LANG GMBH
archenweg 56
innsbruck, 6020
AU  6020
MDR Report Key10395106
Report Number8020045-2020-00017
Device Sequence Number19720809
Product Code DRX
UDI-Device Identifier10861779000274
UDI-Public(01)10861779000274
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K023503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date11/04/2022
Device Model NumberT-601
Device Catalogue NumberSBT601
Device Lot Number191104-0157
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2020
Supplement Date Received by Manufacturer07/28/2020
Initial Report FDA Received Date08/12/2020
Supplement Report FDA Received Date11/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-