| Model Number |
MMT-751LNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423); Battery Problem (2885); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm, scratches on screen and had unexplained no delivery alarms.No harm requiring medical intervention was reported.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but motor error alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to motor error alarm.Motor passed motor test.No charge battery anomaly noted.Device received with cracked battery tube threads, minor scratched lcd window, cracked reservoir tube lip and cracked belt clip slot.The p-cap locks into the reservoir compartment properly noted.(b)(4).
|
| |
|