As the device was being deployed, the physician noted that blood was coming back into the indeflator of the valve delivery system.The indeflator was disconnected, and there was noted to be risk, pulsatile bleed back from the delivery system signifying that the tear in the balloon was large.Upon removing the valve and esheath, there was resistance in the iliac artery and the patients blood pressure dropped, cpr was initiated.With removal of the undeployed valve, it was noted that the external iliac artery had avulsed and come out with the valve as it was stuck on the valve.
|