| Model Number |
MMT-332A |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the reservoir had air bubbles.The customer noticed the air bubbles during therapy.Customer did not allow for insulin to warm to room temperature prior to filling reservoir.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|