• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. G2 EXPRESS FILTER SYSTEM - FEMORAL; VENA CAVA FILTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. G2 EXPRESS FILTER SYSTEM - FEMORAL; VENA CAVA FILTER Back to Search Results
Catalog Number RF400F
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Abdominal Pain (1685)
Event Date 05/21/2015
Event Type  Injury  
Manufacturer Narrative
Manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record review is not required.Investigation summary: the device was not returned for evaluation.Medical records were provided and reviewed.Follow up imaging showed excellent placement of the filter at the level of the renal veins.Approximately six years and three months post filter deployment, patient presented with bilateral upper quadrant pain.Subsequent, computed tomography (ct) revealed that there was small pericardial effusion.An inferior vena cava filter was present.Multiple struts were penetrated through the inferior vena cava wall.Therefore, the investigation is confirmed for perforation of the inferior vena cava (ivc).Based upon the available information, the definitive root cause is unknown.Labeling review: a review of product labeling documents (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, and unit label) showed that the product labeling is adequate.(expiry date: 10/2011).
 
Event Description
It was reported through the litigation process that a vena cava filter was placed in a patient after being diagnosed with deep vein thrombosis/pulmonary embolism.Approximately six years and a month later, a computed tomography(ct) scan demonstrated that the filter struts perforated outside the inferior vena cava wall.The device has not been removed and there were no reported attempts made to retrieve the filter.The current status of the patient is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G2 EXPRESS FILTER SYSTEM - FEMORAL
Type of Device
VENA CAVA FILTER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury NY 12804
Manufacturer Contact
judith ludwig
1415 w. 3rd street
tempe, AZ 85281
4803032689
MDR Report Key10395327
MDR Text Key202514538
Report Number2020394-2020-05357
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080668
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRF400F
Device Lot NumberGFSI2360
Was Device Available for Evaluation? No
Date Manufacturer Received07/17/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age57 YR
-
-