MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1000 MED-EL CONCERT
Medical Device Problem Codes Circuit Failure (1089); Output Problem (3005)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/19/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
The user_s hearing performance using the device was affected.The user has been re-implanted on (b)(6) 2020.
 
Additional Manufacturer Narrative
Conclusions: device investigation confirmed that the stimulator electronics is not working according to specifications.The failure of the electronic circuitry was likely caused by humidity ingress at the detected micro-leaks at the housing's header assembly.Over a certain period of time, humidity will impinge on the electronics and cause damage, potentially leading to complete failure of the circuitry.The investigation findings appear to match the content of the patient report.This is a final report.
 
Event or Problem Description
The recipient's hearing performance using the device was affected.The recipient has been re-implanted on (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key10395629
Report Number9710014-2020-00440
Device Sequence Number2618817
Product Code MCM
UDI-Device Identifier09008737062743
UDI-Public(01)09008737062743
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMI1000 MED-EL CONCERT
Device Catalogue Number07673
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/06/2020
Supplement Date Received by Manufacturer08/06/2020
Initial Report FDA Received Date08/12/2020
Supplement Report FDA Received Date09/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
-
-