| Model Number |
MI1000 MED-EL CONCERT |
| Medical Device Problem Codes |
Circuit Failure (1089); Output Problem (3005)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
04/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The user_s hearing performance using the device was affected.The user has been re-implanted on (b)(6) 2020.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusions: device investigation confirmed that the stimulator electronics is not working according to specifications.The failure of the electronic circuitry was likely caused by humidity ingress at the detected micro-leaks at the housing's header assembly.Over a certain period of time, humidity will impinge on the electronics and cause damage, potentially leading to complete failure of the circuitry.The investigation findings appear to match the content of the patient report.This is a final report.
|
| |
|
Event or Problem Description
|
|
The recipient's hearing performance using the device was affected.The recipient has been re-implanted on (b)(6) 2020.
|
| |
|