Model Number 16-02-80 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no known patient involvement and no patient information has been provided.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report, that a heater-cooler system 3t was found to be contaminated with mycobacterium chimaera.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
H10: through follow up communication livanova learned that the device was cleaned regularly per the instruction for use and that it was placed inside the operating theatre during use with the fan directed away from the patient and at an estimated distance of 2.5 meters from the surgery field.The affected device was received at the manufacturer site for deep disinfection and the reported contamination could be confirmed.
|
|