U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLANCE HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLANCE HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/21/2020
Event Type  malfunction  
Manufacturer Narrative
There was no known patient involvement and no patient information has been provided.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Corrective actions are in progress for this issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report, that a heater-cooler system 3t was found to be contaminated with mycobacterium chimaera.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: through follow up communication livanova learned that the device was cleaned regularly per the instruction for use and that it was placed inside the operating theatre during use with the fan directed away from the patient and at an estimated distance of 2.5 meters from the surgery field.The affected device was received at the manufacturer site for deep disinfection and the reported contamination could be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLANCE
lindberghstr. 25
munich
MDR Report Key10395668
MDR Text Key202591543
Report Number9611109-2020-00457
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number16-02-80
Was the Report Sent to FDA? No
Date Manufacturer Received09/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-