• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 5 GANG 4-WAY NANOCLAVE STOPCOCK W/BASEPLATE, ROTATING LUER; STOPCOCK, I.V. SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. 5 GANG 4-WAY NANOCLAVE STOPCOCK W/BASEPLATE, ROTATING LUER; STOPCOCK, I.V. SET Back to Search Results
Catalog Number AC205
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Underdose (2542)
Date of Event 07/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received.Investigation is not yet complete.
 
Event or Problem Description
The event involves an ac205, 5 gang 4-way nanoclave® stopcock w/baseplate, rotating luer that the customer reported the multiple luer lock connections broke off after a coronary artery bypass graft.The customer mentioned that one of these ports had epinephrine infusing which negatively impacted the patient after the failure of the luer lock and the patient was stabilized.There was patient involvement with an unspecified adverse event and unspecified patient harm along with a delay in therapy; however, no additional information was provided.
 
Additional Manufacturer Narrative
H10: one (1) used list # ac205, 5 gang 4-way nanoclave¿ stopcock w/baseplate, rotating luer; lot # 3553144 was received and visually inspected.As received, all five nanoclaves ports were separated from the manifold body due to insufficient uv adhesive coverage at the bonded connection.The probable cause is due to an error in the manual bonding process during assembly.The device history review (dhr) for lot number 3553144 was reviewed and there were no relevant non-conformances found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5 GANG 4-WAY NANOCLAVE STOPCOCK W/BASEPLATE, ROTATING LUER
Common Device Name
STOPCOCK, I.V. SET
Manufacturer (Section D)
ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
avenida cuarzo no. 250
ensenada, b.cfa. 22790
MX  22790
MDR Report Key10395785
Report Number9617594-2020-00309
Device Sequence Number11403584
Product Code FMG
UDI-Device Identifier00887709080010
UDI-Public(01)00887709080010(17)221001(10)3553144
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K080077
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2022
Device Catalogue NumberAC205
Device Lot Number3553144
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2020
Initial Date Received by Manufacturer 07/20/2020
Supplement Date Received by Manufacturer09/08/2020
Initial Report FDA Received Date08/12/2020
Supplement Report FDA Received Date09/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EPINEPHRINE, MFR UNK; EPINEPHRINE, MFR UNK
Outcome Attributed to Adverse Event Required Intervention;
-
-