| Model Number |
MMT-523RNAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/07/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that they experienced high blood glucose.The customer¿s blood glucose level was 400 mg/dl.The customer was reported that insulin pump had button error alarm, the alarm was not cleared.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|