MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC FUSION COMPACT EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC FUSION COMPACT EAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735602
Device Problems Connection Problem (2900); Device Sensing Problem (2917)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/07/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9735638, serial/lot #: n/a, version: (b)(4). Patient information has been requested. Surgeon information has been requested. System checkout is pending case generation and follow up on if troubleshooting resolved the issue. If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation system being used during a procedure. It was reported that the emitter was not picking up the patient tracker. With a new patient tracker, the emitter would pick it up briefly and then go red status again. The entire emitter detail pane was red, including the axiem box and emitter. Troubleshooting was attempted to resolve the issue. The axiem box stated that it was not communicating and the emitter was unable to read the port. The emitter was disconnected from the axiem, but this did not resolve the issue. Navigation was aborted. There is no further information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFUSION COMPACT
Type of DeviceEAR, NOSE AND THROAT STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key10396188
MDR Text Key202601213
Report Number1723170-2020-02150
Device Sequence Number1
Product Code PGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received08/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9735602
Device Catalogue Number9735602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received11/02/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-