Model Number 10300
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Event or Problem Description
A customer in (b)(6) notified biomérieux of misidentifications of listeria monocytogenes when testing two (2) strains with api® listeria 10strips+10media (ref 10300, lot 1007942830).The listeria monocytogenes strains included one (1) isolated from food samples and one (1) internal l.Monocytogenes qc strain (atcc® 13932¿).Api obtained listeria welshimeri and listeria innocua as significant taxon for the strain isolated from food samples and listeria welshimeri as significant taxon for the qc strain.The expected identification of listeria monocytogenes was confirmed by a reference lab with pcr and sero-type identification.There is no patient involved as food and qc strains are impacted.A biomérieux internal investigation will be initiated.
 
Additional Manufacturer Narrative
This report was initially submitted following notification from a customer in switzerland regarding misidentifications of listeria monocytogenes when testing two (2) strains with api® listeria 10strips+10media (ref 10300, lot 1007942830).The listeria monocytogenes strains included one (1) isolated from food samples and one (1) internal l.Monocytogenes qc strain (atcc® 13932¿).The investigation included batch record analysis.There were no nonconformities observed related to the subject lot.Quality control results were in compliance with specifications for batch release.Internal testing was performed on api® listeria 10strips+10media lot 1007942830 (related to the complaint) and lot 1007999710 (used as a reference lot).Three (3) strains mentioned in the package insert api listeria 07887 t, and one (1) strain from r&d (atcc® 13932¿) were tested on the lots.There were no performance issues or identification errors observed during testing.The customers misidentification was not reproduced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
API LISTERIA 10STRIPS+10MEDIA
Common Device Name
API® LISTERIA 10STRIPS+10MEDIA
Manufacturer (Section D)
BIOMERIEUX, SA
3 route de port michaud
la balme 38390
FR  38390
MDR Report Key10398218
Report Number9615754-2020-00135
Device Sequence Number2704905
Product Code JTO
UDI-Device Identifier03573026035587
UDI-Public03573026035587
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
C1 EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2021
Device Model Number10300
Device Catalogue Number10300
Device Lot Number1007942830
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2020
Supplement Date Received by Manufacturer08/04/2020
Initial Report FDA Received Date08/12/2020
Supplement Report FDA Received Date08/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-