| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter in two pieces.The balloon was tightly folded.There was blood in the inner lumen/shaft.Analysis of the tip, balloon, inner/outer shaft and hypotube included microscopic and visual inspection.Inspection revealed a kink in the hypotube located 61.5cm form the tip and a complete fracture in the hypotube located 72.5cm from the tip.The fracture faces of the separated were ovalized, as if kinked prior to being separated.Inspection of the rest of the device found no other damage or defect.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 22jul2020.It was reported that shaft kink occurred.The 93% stenosed, 13mmx2.5mm target lesion was located in the moderately tortuous and calcified left anterior descending artery.A 2.5mm x 12mm quantum maverick balloon catheter was advanced; however, the balloon delivery shaft was kinked.The procedure was completed with another of same device.No patient complications were reported and the patient was stable.However, returned device analysis revealed shaft separation.
|
| |
|