| Model Number |
00884450333327 |
| Medical Device Problem Codes |
Fracture (1260); Material Fragmentation (1261)
|
| Health Effect - Clinical Codes |
Needle Stick/Puncture (2462); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
08/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that the biopsy device broke off and the tip of the coaxial introducer remained within the patient's body during a bone biopsy.The separated portion was approx.1 cm in length and it remained in the body.Secondary procedure is required for the remainder, although it has not been scheduled yet.The condition of the patient has not been changed after the event, except for the tip detachment itself.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed; however, the root cause was attributed to clinical use contraindicated in the instructions for use (ifu) of the device.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|