EXACTECH, INC. TRULIANT TIB IMP PSC INSERT SZ 4.5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
|
Back to Search Results |
|
| Model Number |
02-022-44-4511 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, a (b)(6) y/o, male patient with a bmi of 35.9, underwent implantation of a insert tibial truliant psc sz4.5 11mm.On (b)(6) 2019, approximately 11 months post implant, the patient underwent revision due to instability.
|
| |
|
Additional Manufacturer Narrative
|
|
After further review of additional information received the following sections g3, g4, g7, h2 and h3 have been updated accordingly.(h3) upon review of all available information, there is no evidence to suggest that a design or manufacturing issue caused or contributed to this event.The reported revision was likely the result of underlying patient factors, which led to instability of the joint.Per ifu# 700-096-004 instability is a known complication following total knee replacement surgery and is multifactorial in nature.Instability is listed as a potential adverse event associated with total joint replacement surgery in the product ifu (700-096-004).As noted in why are total knee arthroplasties failing today has anything changed after 10 years? (sharkey, p.L.(2014s, september).The journal of arthroplasty, 29), instability is in the top five reasons for that account for approximately 85% of total knee revisions (7.35% of revisions are for instability).
|
| |
|
Search Alerts/Recalls
|
|
|