STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6020-0335
Medical Device Problem Codes Corroded (1131); Degraded (1153); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Injury (2348); Hip Fracture (2349); Osteolysis (2377); Reaction (2414); Metal Related Pathology (4530)
Date of Event 07/21/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported that the patient's left hip was revised due to metallosis and elevated metal ions.An lfit v40 head and poly liner were revised to a ceramic on poly bearing.Rep confirmed there are no allegations against the revised liner, and that no further information will be released by the hospital or surgeon.
 
Event or Problem Description
It was reported that the patient's left hip was revised due to metallosis and elevated metal ions.An lfit v40 head and poly liner were revised to a ceramic on poly bearing.Rep confirmed there are no allegations against the revised liner, and that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level, altr, corrosion, periprosthetic fracture, metallosis and osteolysis involving an accolade stem was reported.The event was confirmed only for altr, corrosion, metallosis based on medical review.Method & results: -product evaluation and results: not performed as no items were returned.-clinician review: a review of the provided medical records by a clinical consultant stated the following comment:(b)(6) 2020: stryker pi report (b)(6) 2013: implant sheet.Trident psl ha cluster shell 52-e, 25mm, 25mm, 20mm, screws 6.5 mm torx, 36-e 100 polyethylene liner, #3 accolade tmzf v40 1320 stem, lfit anatomic v40 36 +5mm head.(b)(6) 2020: operative note: revision left tha.Failed tha with metal taper corrosion, chronic displaced greater trochanteric fracture with extensive abductor necrosis and pseudotumor.Specimens sent to path including pseudotumor.Implants 52mm 00 polyethylene, 36mm +7.5 mm biolox ceramic head with titanium sleeve.Had metal levels and mri c/w metal taper corrosion, pseudotumor, osteolysis induced fracture of trochanter, chronic, and abductor necrosis from metallosis.Had cardiac procedure preop for optimization.Evidence of metallosis in joint and 70% of abductors involved.Osteolysis around the fractured troch.Extensive corrosion on the taper and head when removed.Cup and stem stable.Considered mdm and constraint but went with poly due to metallosis.Capsule and troch repaired with ethibond suture.Confirmation: revision surgery for metallosis and taper corrosion in the setting of an lfit v40 head and tmzf accolade stem could be confirmed.The increased metal levels could not be confirmed.Root cause: the root cause of the revision was trunnion corrosion and metallosis leading to abductor necrosis and pseudotumor.The root cause of the trunnion corrosion was likely a v40-lfit trunnion issue (this could be confirmed with lot numbers).The metal levels were not confirmed but if elevated would have the same root cause.-product history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: it was reported that patient's left hip was revised due to abnormal ion level, altr, periprosthetic fracture, corrosion, metallosis and osteolysis involving a metal head was reported.The event was confirmed only for altr, corrosion, metallosis based on medical review.The exact cause of the event could not be determined because insufficient information was provided.Additional information including return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #3
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10400522
Report Number0002249697-2020-01654
Device Sequence Number17108640
Product Code LPH
UDI-Device Identifier04546540510556
UDI-Public04546540510556
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2018
Device Model Number6020-0335
Device Catalogue Number6020-0335
Device Lot Number44345802
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/21/2020
Supplement Date Received by Manufacturer09/17/2021
Initial Report FDA Received Date08/13/2020
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age88 YR
-
-