| Model Number |
MMT-723NAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had instances of no delivery and motor errors alarm.No harm requiring medical intervention was reported.The customer was assisted with troubleshooting.The drive support cap was flushed.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but motor error alarm during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery alarm due to motor error alarm.Motor passed motor test.
|
| |
|