• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN. Back to Search Results
Model Number 52969
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/23/2020
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical inc.Is currently investigating the reported condition.
 
Event Description
Loose conn.On probe.S&r had to replace diaphgram.Order: 94659.Ref: e-complaint-(b)(4).Leep system 1000 esu gen 52969 e-complaint-(b)(4).
 
Event Description
Loose conn.On probe.S&r had to replace diaphgram.Order: (b)(4).Ref: (b)(4).1216677-2020-00175 leep system 1000 esu gen 52969 (b)(4).
 
Manufacturer Narrative
Investigation.X- inspect returned samples.Analysis and findings: (b)(4).Distribution history: this complaint unit was manufactured at csi on 08/05/2010 under wo# (b)(4) and shipped on 8/27/2010.Manufacturing record review: a review of the device history record could not be performed as the record could not be located at this time.However, at the time of manufacture, records from each unit are reviewed to ensure that product meet all specifications.Should the device history record be located, this complaint will be amended accordingly.Service history record: no additional service history records found for this unit.Historical complaint review: a review of the2-year complaint history showed similar reported complaint conditions.Product receipt: the complaint unit was returned on a repair.However, based on log 94659, this unit was at csi on 7/29/2020.Visual evaluation: visual examination of the complaint unit revealed physical damage.Functional evaluation: unit was not functioning properly.Root cause: a loose handpiece socket connection is due to wear and tear.The diaphragm was changed out due to a previous finding where the material degraded to the point where it no longer functioned as intended and prevents activation of a unit.This unit's diaphragm was in working order.However, the corrective action for this issue requires returned units be updated to the new material and applicable on this unit.Correction and/or corrective action / *preventative action activity the unit was repaired, tested to specifications and returned tot he customer.Coopersurgical will continue to monitor this complaint condition for trends.No further corrective action is necessary.Corrective actions relevant to the updated diaphragm are as follows; coopersurgical service and repair replaced the diaphragm on the unit, tested it and returned it to the customer.Engineering has successfully tested a replacement material made of silicone for use in repairs going forward, eng-test-10341-r.The ifu was also updated to add a safety check via ecn-20444.A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.The repaired unit will have been repaired with this new silicone material.No further training required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN.
Type of Device
LEEP SYSTEM 1000 ESU GEN.
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull CT 06611
Manufacturer (Section G)
COOPERSURGICAL, INC.
75 corporate drive
trumbull CT 06611
Manufacturer Contact
michael marone
50 corporate drive
trumbull, CT 06611
4752651665
MDR Report Key10401105
MDR Text Key209030507
Report Number1216677-2020-00175
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K952483DISCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial,Followup
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number52969
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/05/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-