• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMERGE; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7134
Medical Device Problem Codes Break (1069); Entrapment of Device (1212)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/22/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheter entrapment and shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was advanced for dilation; however, the balloon became stuck in the bmw guide wire.Both devices were removed together from the patient's body and when tried to remove the balloon from the wire, the balloon shaft broke.The procedure was completed with a different device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 106.2cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The inner was stretched down starting 54mm from the tip and extending 5mm, and at 36mm extending 15mm.The inner shaft was buckled for 9mm starting 9mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that catheter entrapment and shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was advanced for dilation; however, the balloon became stuck in the bmw guide wire.Both devices were removed together from the patient's body and when tried to remove the balloon from the wire, the balloon shaft broke.The procedure was completed with a different device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMERGE
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10401177
Report Number2134265-2020-10259
Device Sequence Number3093683
Product Code LOX
UDI-Device Identifier08714729806349
UDI-Public08714729806349
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K113220
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/24/2022
Device Model Number7134
Device Catalogue Number7134
Device Lot Number0024838928
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer09/02/2020
Initial Report FDA Received Date08/13/2020
Supplement Report FDA Received Date09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-