| Model Number |
7134 |
| Medical Device Problem Codes |
Break (1069); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment and shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was advanced for dilation; however, the balloon became stuck in the bmw guide wire.Both devices were removed together from the patient's body and when tried to remove the balloon from the wire, the balloon shaft broke.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 106.2cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was loosely folded.Microscopic inspection revealed tip damage.The inner was stretched down starting 54mm from the tip and extending 5mm, and at 36mm extending 15mm.The inner shaft was buckled for 9mm starting 9mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment and shaft break occurred.A 3.00mm x 20mm emerge balloon catheter was advanced for dilation; however, the balloon became stuck in the bmw guide wire.Both devices were removed together from the patient's body and when tried to remove the balloon from the wire, the balloon shaft broke.The procedure was completed with a different device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|