| Model Number |
20005D |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Keratitis (1944)
|
| Date of Event |
07/17/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The system was evaluated by a field service engineer.The field service found no issues with the unit.A field service checklist was performed.The unit complied with all factory settings.A record review was performed.A product deficiency review was performed and there is no product deficiency identified.A document, service history, and trending was reviewed.There is not a recognizable adverse trend.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.A labeling review was conducted; the operator manual for the system was reviewed and found to include adequate instructions for use, warnings and operational errors.The review of the device history record (dhr) for femtosecond laser system showed that there were no issues or non-conformities.The system and its components met all specifications prior to being released.Manufacturing has been ruled out as a potential cause for the reported issue.Based on the investigation results, no corrective action has been issued.Based on the investigation results there is no indication of a product quality deficiency.Johnson & johnson surgical vision will continue to monitor this type of complaints.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that two patients presented with diffuse lamellar keratitis (dlk and both patients were treated with topical steroids.In addition, it was reported that one patient was a female and one was a male.One had stage 2 dlk and the other had stage 3 dlk.One was prescribed medrol pack and had a flap lift and rinse performed. the additional information that was provided, did not indicate which of the patients had the surgical intervention ( flap lift and rinse) or medrol prescribed; nor what stage of dlk for each gender.Additional follow attempts were made with the account to obtain specific patient information however, no response was received from the account. this report is for female patient #1 using the femto laser.A separate report will be filed for female patient #1 using the excimer laser. patient #2 was submitted under manufacturing report # 3006695864-2020-00383 and 3006695864-2020-00384.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|