MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724
Device Problems Disconnection (1171); Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 07/22/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the em2400 valve set leaked where the tubing connected to the hard plastic of the valve body.It was further reported that when attempting to redo the whole setup, the tubing "split at the joint".This was identified during setup prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added d10, h3, h4 and h6.H4: device manufactured between march 06, 2018 to march 07, 2018.H10: one (1) actual and one (1) companion devices were received for evaluation.A visual inspection was done and observed the actual sample was received with damaged tubing and companion sample with misaligned tubing.No functional testing was performed for this complaint.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10402640
MDR Text Key202754389
Report Number1416980-2020-04942
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477183
UDI-Public(01)00085412477183
Combination Product (y/n)Y
PMA/PMN Number
K002705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2020
Device Catalogue NumberH938724
Device Lot Number60116213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2020
Date Manufacturer Received09/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-