• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGITIE; SUTURE, ABSORBABLE, NATURAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGITIE; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Model Number EL-21-L
Medical Device Problem Codes Break (1069); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during an appendectomy, on endoscopic suturing, the shaft that connects the suture was broken out of box.Another device was used to complete the case.There was no patient injury.
 
Event or Problem Description
According to the reporter, during an appendectomy, on endoscopic suturing, the shaft that connects the suture was broken upon opening the package.Another device was used to complete the case.There was no patient injury.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the returned product noted one suture, two nylon shaft, and one delivery tube.As for packaging, a blister and a stapled foil pouch was received.A nylon shaft was received with the proximal end missing.This nylon shaft was inside a delivery tube.The remaining nylon shaft was received with the proximal end broken.The suture was still going through the nylon shaft and the loop was partially closed while the 5 inches of suture separated the proximal end and the rest of the nylon shaft.The nylon shaft was received secure parked inn the blister.It was reported that the needle unexpectedly separated from the thread.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device histo ry records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGITIE
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key10402815
Report Number9612501-2020-01155
Device Sequence Number14187517
Product Code GAL
UDI-Device Identifier10884521099630
UDI-Public10884521099630
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K904588
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberEL-21-L
Device Catalogue NumberEL-21-L
Device Lot NumberJ9F2637Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2020
Initial Date Received by Manufacturer 07/28/2020
Supplement Date Received by Manufacturer12/07/2020
Initial Report FDA Received Date08/13/2020
Supplement Report FDA Received Date12/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-