OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 07/17/2020
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was requested to go to the user facility to evaluate the device as no guarantee could be made on maintaining a positive channel pressure on the device with the broken connector.To date, the fse visit has not been scheduled, pending customer po.A supplemental report will be submitted if new information becomes available and upon completion of the evaluation of the device.
 
Event Description
The user facility reported that the oer-pro has a broken scope connector, the grey air/water channel connector is broken off.There was no patient involvement.No additional information was provided.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: "check the following for each connector.The connector should be fixed firmly.The o-rings should be free of abnormalities such as cracks, tears, or dents¿.The root cause could not be conclusively determined.Probable causes that could lead to the reported event include that the connector was attached forcibly, accumulating stress which led to the breakage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10403170
MDR Text Key202803473
Report Number8010047-2020-05469
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 11/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-