• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 3 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-030
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 07/31/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had a right tha performed on (b)(6) 2020.The patient was doing great until she tripped on carpet at home.From the xrays she fx her femur and has a periprosthetic fracture.The surgeon wanted to remove the trilock stem and head, and put in a corail revision stem while cabling the femur.The surgeon liked the final result in regards to leg length, offset, and new stability with the corail revision stem.Doi: (b)(6) 2020, dor: (b)(6) 2020, right side.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.A review of a singular x-ray image does find sufficient evidence to confirm the reported bone fracture.No device error is not however identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 3 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10403419
Report Number1818910-2020-18092
Device Sequence Number19630899
Product Code KWA
UDI-Device Identifier10603295000747
UDI-Public10603295000747
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-04-030
Device Catalogue Number101204030
Device Lot NumberJ62X47
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/31/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date08/13/2020
Supplement Report FDA Received Date09/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 32 MM +5; DELTA CER HEAD 12/14 32 MM +5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-