| Model Number |
ZCB00 |
| Medical Device Problem Code |
Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Age/date of birth: unknown/ not provided.Sex/gender: unknown/ not provided.If implanted, give date: n/a (not applicable).The intraocular lens was not implanted.If explanted, give date: n/a (not applicable).The intraocular lens was not implanted; therefore, not explanted.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
The account reported that they had difficulties loading the intraocular lens (iol) and tried three (3) times to load the lens.It was indicated that the doctor always makes small incision.He attempted to insert the lens and enlarged the incision.A new replacement lens was inserted without difficulty.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the product was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated, and no deviation or nonconformance was found during process related to the complaint issue reported.The product was manufactured and released according to specifications.A search in complaint system revealed that no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|