• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TCM COMBIM MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TCM COMBIM MODULE Back to Search Results
Model Number 903-111
Medical Device Problem Codes Insufficient Device Problem Information (3190); Device Handling Problem (3265)
Health Effect - Clinical Codes Burn, Thermal (2530); Partial thickness (Second Degree) Burn (2694)
Date of Event 07/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Please note that the tcm combim monitoring system consist of a tcm base unit (item number: 391-880) and the combim module (903-111).The combim module is referenced as this is the affected model in this report, however please note that the serial number and udi number provided , is associated to the tcm base unit, as this is the only serial number currently known from the complaint.Further, please not that the manufacturing date in field is associated to the base unit (391-880).In the report the serial number and associated udi number has been stated for the base unit with serial number: (b)(4).As the customer is not sure whether the base unit involved in the incident was the one with serial number (b)(4), the udi for n006 is included here: (b)(4).
 
Event or Problem Description
A set of twins 32 weeks of age were monitored by two tcm combim monitors (serial numbers (b)(4)).Twin (a) appeared to have a bruise on her outer thigh.The customer also states that this patient has similar deep red marks bilaterally as well on her back; up by her shoulders.The other twin (b) had no concerns or issues at this time.The customer reached out for a wound consult who recommended to use radiogel sheets every 2 to 3 days and follow up.Both monitors were taken off the patients and no longer in service.The customer states she does not know which monitor was on each twin, but the monitor ending in n005 was set to 44 degrees c, site time set to 6 hours, and the co2 alarm is set to off.The monitor ending in n006 was set to 42 degrees.Have not confirmed if the co2 alarm was on or off or the length of time set on the site time.Both monitors time and date was incorrect.The sensor placement on twin a was alternating between outer legs beginning at 1030 am on 7/15 to 7/16.On 7/17 the sensor placement changed to the back for approx.For 6 hours until it was removed off of twin a on the morning of 7/17.At this time both monitors were taken off of both babies.
 
Additional Manufacturer Narrative
By a mistake this additional information was not included in the mdr 3002807968-2020-00034 follow-up #1 report.Please see the additional information below: the sensor placement on the patient was alternating between outer legs beginning at 10:30 am on 7/15 to 7/16.On 7/17 the sensor placement changed to the back for approx.For 6 hours until it was removed off of the patient on the morning of 7/17.According to tcm tosca/combim operator's manual (201811i9, page 6-3) the recommended maximum sensor application time is 4 hours for a maximum sensor core temperature of 44°c for neonates (excluding preterm infants) and adults.The root cause is determined as too long exposure to a sensor temperature of 44°c.All h6 codes have been updated in this mdr follow-up #2 report to follow the new layout of the medwatch form 3500a.It has not been possible to fill out the 'health effect -impact code' in h6 in the esubmitter system.The code should be "appropriate term/code not available".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TCM COMBIM MODULE
Common Device Name
TCM COMBIM MODULE
MDR Report Key10406799
Report Number3002807968-2020-00034
Device Sequence Number11401117
Product Code LKD
UDI-Device Identifier05700699031112
UDI-Public(01)05700699031112
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K093154
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number903-111
Device Catalogue Number903-111
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/17/2020
Supplement Date Received by Manufacturer08/17/2020
08/17/2020
Initial Report FDA Received Date08/14/2020
Supplement Report FDA Received Date09/08/2020
09/08/2020
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age32 WK
-
-