Model Number 102096-67A |
Device Problem
Failure to Sense (1559)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
|
Event Description
|
On (b)(6) 2020,senseonics was made aware of an incident where user received an early sensor replacement alert resulting in an early sensor removal.
|
|
Manufacturer Narrative
|
The sensor was investigated and the issue was confirmed during qc testing in-house.Intermittent sensor led failures were detected during this test.The investigation shows the system correctly disabled the sensor due to performance failure and the system's self-test functions are working normally.D8 was this device serviced by a third party? no.D9 device available for evaluation? yes, device received on 10/15/2020.H3.Device evaluated by manufacturer? yes.H6.Health effect - clinical code updated to 4582.H6.Health effect -impact code updated to 2199.H6.Medical device problem code updated to 1559.H6.Component code updated to 510.H6.Type of investigation updated to 4111 and 10.H6.Investigation findings updated to 4203.H6.Investigation conclusions updated to 4307.
|
|