| Model Number |
5177801400 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release and no trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Once our evaluation is complete, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
According to the available information, explanted penile prosthesis due to: "displacement/infection".
|
| |
|
Event or Problem Description
|
|
This follow-up mdr is created to document the additional event information received for record #601873.According to the available information the inflatable penile prosthesis was implanted on (b)(6) 2008 and removed on (b)(6) 2020 due to displacement/infection.Additional information received from the pathology coordinator indicated: ¿explanted penile prosthesis due to: ¿displacement/infection¿.The device was not received for evaluation.As examination of the device may not conclusively confirm or disprove the report of infection quality accepts the physician¿s observations as to the reason for surgical intervention.The review of the device lot history records indicates this lot passed sterility testing prior to being released.The device was sterile prior to the device packaging being opened and that the infection originated from source(s) other than the device.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database revealed no trends for this lot.Review of nonconforming reports revealed no nonconformances for this lot.No capas are associated with this lot.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information, the reported complaint is identified in the risk management documentation and reviewed routinely with management to monitor complaint trends as part of post market surveillance.No corrective action is required at this time.
|
| |
|