AVANOS MEDICAL INC. MIC SAFETY PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PEG KIT WITH ENFIT CONNECTORS; DH EF PEG INITIAL PLACEMENT PRODUCTS Back to Search Results
Model Number 8170-20
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Laceration(s) of Esophagus (2398)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of 14-aug-2020 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported a patient came in for a peg tube replacement.The old peg tube was not able to be removed via the insertion site, due to rigidity of the disk.It had to be removed via scope, up and through the esophagus, causing an esophageal injury seen via computed tomography (ct) of the head/neck.The patient did not require intervention in the recovery area.Additional information received 29-jul-2020 stated the patient did not require any surgical intervention, only antibiotics, and has since been discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC SAFETY PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PEG KIT WITH ENFIT CONNECTORS
Common Device Name
DH EF PEG INITIAL PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key10410662
Report Number3006646024-2020-00031
Device Sequence Number3092563
Product Code KNT
UDI-Device Identifier00350770446004
UDI-Public00350770446004
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K924065
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number8170-20
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2020
Initial Report FDA Received Date08/14/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient Weight51
-
-