DEPUY IRELAND - 9616671 RECLAIM DISASMBLY THREADED CAP; RECLAIM INSTRUMENTS : EXTRACTION INSTRUMENTS Back to Search Results
Model Number 2975-00-710
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Date of Event 08/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the instruments broke while trying to disengage proximal body.There was an (1) one hour of surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument found the alleged complaint unconfirmed but found a different failure upon inspection of the instrument.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: examination of the returned instrument cannot confirm the complaint that instrument is broken.However, visual exam found the threaded cap is cross threaded on the returned reclaim dismbly trq shft 95mm and cannot be disassembled.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DISASMBLY THREADED CAP
Common Device Name
RECLAIM INSTRUMENTS : EXTRACTION INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10410764
Report Number1818910-2020-18188
Device Sequence Number2552794
Product Code LXH
UDI-Device Identifier10603295156079
UDI-Public10603295156079
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2975-00-710
Device Catalogue Number297500710
Device Lot NumberJ0911
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Initial Date Received by Manufacturer 08/04/2020
Supplement Date Received by Manufacturer08/14/2020
09/16/2020
09/16/2020
Initial Report FDA Received Date08/14/2020
Supplement Report FDA Received Date08/25/2020
09/17/2020
09/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-