| Model Number |
305905 |
| Medical Device Problem Code |
Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that syringe 3ml ll w/ndl sftygld 23x1 rb had a safety mechanism failure.This was discovered before use.The following information was provided by the initial reporter: safety mechanism can not work well.
|
| |
|
Additional Manufacturer Narrative
|
|
D.10 device available for eval yes, d.10 returned to manufacturer on: 2020-08-17.H.6.Investigation summary one photo and one 3ml safetyglide combo in an opened blister from batch 9287062 (p/n 305905) was received and evaluated.It was observed the ribs of the hub were damaged with small plastic fragments protruding outwards and the safety mechanism was disconnected, which was rejectable per product specification.Potential root cause for the broken safety mechanism defect is associated with the needle manufacturing process.The samples will be forwarded to the needle manufacturing site for evaluation and potential root cause determination.Bd columbus was able to identify the root cause for the defect occurred on the safetyglide assembly line.The plastic hub is placed under the cannulator then the needle is positioned and assembled to the plastic hub adding the epoxy to fix it; after that, a plastic safety mechanism is assembled, then at the end the plastic shield is assembled to the part.In this case, the needle assembly mechanism was not properly assembled to the safety mechanism inducing the symptom reported.Based on the investigation and the analysis of the sample the symptom reported by the customer is verified.No corrective actions are required at this time.Bd will continue to monitor this lot and failure mode through monthly trending meetings.
|
| |
|
Event or Problem Description
|
|
It was reported that syringe 3ml ll w/ndl sftygld 23x1 rb had a safety mechanism failure.This was discovered before use.The following information was provided by the initial reporter: safety mechanism can not work well.
|
| |
|