Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problems
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
Event Date 07/20/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a patient underwent an initial knee arthroplasty on (b)(6) 2020.During the procedure the 3.2 drill broke while in use during a total knee arthroplasty.The tip of the drill was stuck in the patients femur.An inter-op x-ray was taken to locate the exact location.The surgeon attempted to retrieve the broken piece.The attempt was unsuccessful and the decision was made to leave it.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of x-rays by a third party hcp noting a broken end of drill bit seen in the distal femur.Bone quality appears normal.Dhr was unable to be reviewed as the lot number for the device is unknown.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|