Model Number N/A
Device Problem Fracture (1260)
Patient Problems Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Event Date 07/20/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient underwent an initial knee arthroplasty on (b)(6) 2020.During the procedure the 3.2 drill broke while in use during a total knee arthroplasty.The tip of the drill was stuck in the patients femur.An inter-op x-ray was taken to locate the exact location.The surgeon attempted to retrieve the broken piece.The attempt was unsuccessful and the decision was made to leave it.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of x-rays by a third party hcp noting a broken end of drill bit seen in the distal femur.Bone quality appears normal.Dhr was unable to be reviewed as the lot number for the device is unknown.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW DRILL 3.2 MM DIAMETER
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key10413157
MDR Text Key203106433
Report Number0001822565-2020-02959
Device Sequence Number1
Product Code HTW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00512008500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/22/2020
Initial Date FDA Received08/17/2020
Supplement Dates Manufacturer Received12/15/2020
Supplement Dates FDA Received12/16/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age68 YR
-
-